Horiba Scientific FluoroMax-4 Operation Manual page 42

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
This is an uncalibrated FluoroMax
467 nm, but here appears near 480 nm.
7
Calibrate the excitation monochromator, if re-
quired.
a
b
c
d
e
Structure ~ 450 nm
Note:
Your lamp scan may appear different, depending on the
instrument configuration.
Expand the plot by clicking the Expand button.
Click the cursor button
Click on the graph near the peak, to place the cursor on the graph.
Using the left and right arrows on the keyboard, move the cursor to the
top of the peak.
Read the x-value of this plot: this is the wavelength of the peak.
Calibration peak at 467 nm
Broadband
®
lamp scan. The main peak ought to be at
to start the Cursor function.
3-6
System Operation