Lamp Replacement - Horiba Scientific FluoroMax-4 Operation Manual

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)

Lamp replacement

The xenon flash lamp typically has a half-intensity life of at least 10 million flashes.
Follow the procedure below for replacement and alignment.
Warning:
High voltage exists within the FluoroMax
To avoid fatal shocks, before removing the lamp cover,
unplug the FluoroMax
least 5 min while the internal capacitors discharge.
Never operate the lamp with the cover removed!
Caution:
Intense ultraviolet, visible, or infrared light may
be present when the instrument is open, so wear eye-
and skin-protection, such as light-blocking goggles and
light-blocking clothing.
Warning:
Xenon lamps are an explosion hazard. Be
sure that the power is off, and all AC (mains) power is
disconnected from the system. Read and follow all the
cautions below:
9-11
®
FluoroMax
-4P Phosphorimeter Operation
®
-4P's power cord, and wait at
®
-4P.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fluoromax-4p

Table of Contents