Horiba Scientific FluoroMax-4 Operation Manual page 43

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
Cursor's wave-
length in nm
f
g
This example shows the peak actually at 477 nm, which is 10 nm too
high. Therefore we must recalibrate the monochromator.
Click the Previous Experiment button
The Experiment Setup window appears.
Click the RTC button on the lower right.
The Real Time Control window opens:
Cursor
.
3-7
System Operation

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fluoromax-4p

Table of Contents