Turning On The Detector For The First Time - Thermo Scientific Accela PDA Hardware Manual

Table of Contents

Advertisement

2
Installation

Turning On the Detector for the First Time

Turning On the Detector for the First Time
24
Accela PDA Detector Hardware Manual
To turn on the detector for the first time
1. Place the power switch at the front of the unit in the Off position (released or out
position).
2. Ensure that the PDA detector is properly connected to the data system computer.
3. Ensure that the data system lists the appropriate configuration settings for the PDA
detector.
For information on instrument configuration, see
4. Check the fuse size (see
You can operate the PDA detector at 100/115V or 230V. However, you must ensure that
the appropriate fuses are installed (see
Figure 19. View of the power supply module and the power line label
Table 2. Fuses
Line power voltage
100/115V
230V
5. Attach the power cord to the power entry module on the back panel of the PDA detector
and connect it to the power source.
6. Turn the power on by pushing the power button in to engage it.
The Power LED turns amber as it downloads the operational file, and then turns solid
green. If it does not light at all, see
"Replacing the Fuses"
on
Figure 19
and
Fuse size
T3.15 A
T1.6 A
Chapter 7, "Troubleshooting."
Chapter 3, "Configuration."
page
95).
Table
2).
USE ONLY WITH
250 V FUSES
DISCONNECT POWER
BEFORE REPLACING FUSES