Installing the LightPipe Flowcell

Figure 13. Lightpipe flowcell access cover
Thermo Scientific
The LightPipe flowcell is packed in a small, separate box within the detector shipping carton.
This small box contains the LightPipe flowcell (with a protective cap on each end) and a
plastic bag containing the inlet and outlet tubing and fingertight fittings. The part numbers
for these items are listed in
CAUTION Use only PEEK fittings to connect tubing to the flowcell. The appropriate
PEEK fittings are included in the LightPipe Flowcell Accessory Kit.
To install the LightPipe flowcell
1. Open the front doors of the detector.
2. Unscrew the captive screw that secures the LightPipe flowcell access cover to the front
panel of the detector and pull the cover off (see
LightPipe flowcell
access cover
Table 6
on
page
103.
Captive screw
Slot for
outlet tubing
STANDARD FILTER WHEEL
1 : OPEN
2 : HOLMIUM OXIDE FILTER
Slot for inlet tubing
Installing the LightPipe Flowcell
Figure
13).
Accela PDA Detector Hardware Manual
2
Installation
17