Installing The Detector - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Variable wavelength detectors
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

3.5.2 Installing the Detector

After the USB software for the detector has been installed (→ page 41), install and
configure the detector in Chromeleon:
1. Start the Chromeleon Server Monitor (→ page 41) and the Chromeleon server if they
are not yet running.
2. Start the Chromeleon Server Configuration program by clicking Start on the taskbar.
Point to Programs (or All Programs, depending on the operating system), point to
Chromeleon, and then click Server Configuration.
3. If necessary, click the plus sign next to the server name to display the items
underneath.
4. Select the timebase to which the detector will be assigned, or create a new timebase
(on the Edit menu, click Add Timebase).
5. Open the Add device to timebase dialog box. To do so, click Add Device on the Edit
menu or right-click the timebase and click Add Device on the menu.
6. On the Manufacturers list, click Dionex HPLC: UltiMate 3000 and on the Devices
list, click VWD-3x00(RS) Detector.
7. The configuration pages for the detector are opened. On each page, verify that the
settings are correct and select additional settings if needed. For a description of the
pages, see section 3.5.3.1 (→ page 44).
8. On the File menu, click Save Installation and then close the Server Configuration
program.
Operating Instructions
VWD-3100 and VWD-3400RS Variable Wavelength Detectors
UltiMate 3000 Series:
Page 43

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vwd-3100Vwd-3400rs

Table of Contents