5.2

Installing the Detector

Charged Aerosol Detectors Corona Veo (RS) Operating Manual
The detector is installed and set up by a Thermo Fisher Scientific service
engineer. The service engineer checks that the installation is correct and
that the detector operates as specified. The engineer also demonstrates
the basic operation and main features.
If personnel other than a Thermo Fisher Scientific service engineer
installs the detector, follow the steps below.
1. Pay attention to the safety guidelines and observe all site
requirements.
For the safety guidelines when installing the detector, see
section 5.1, page
52. For the site requirements, see
page
55.
2. Set up the detector hardware. See
3. Set up the flow connections. See
4. Turn on the detector. See
Adjust the display brightness and date and time as required.
5. Installation in the Chromeleon software (optional)
When the display shows the Main Menu after the self-test, set up
the detector in the software. See
TIP Before turning on the power to the detector for the first time, verify
that the chromatography software is installed on the data system
computer. When the power is turned on, the required USB drivers are
automatically found and the Windows™ operating system can detect the
device.
6. Set the nebulizer gas pressure. See
7. Recommended:
Perform Instrument Installation Qualification.
In Chromeleon, a wizard is available to guide you through the
qualification process:
On the Chromeleon 7 Console: Click Tools > Instrument
Qualification > Installation Qualification.
section 5.5, page
section 5.6, page
section 5.7, page
98.
section 6.8.1, page
section 6.5.2, page
5
Installation
section 5.3,
67.
81.
148.
129.
Page 53