Chapter 7 Handling and Control (iCanTM Touchscreen Controller)
The start process is indicated by an audible signal. During the heat-up phase, the displays show
dots (...) instead of values:
• Temperature display [2]
• CO
display [1]
2
• O
display [3]
2
After the heat-up phase has been completed, the control loop actual values are indicated.
Note
Exposure to CO
gas:
2
During the 5-minute heat-up phase of the O
and the CO
control and is disabled.
2

7.8. Event Actions for Settings

Key response:
Pressing a key can increase or reduce a value gradually:
• When the – key [4] or the + key [6] is kept depressed, the system switches to quick run,
• After the key has been kept depressed for more than 3 seconds, the quick run speed increases.
Note
Saving the settings:
To save changed values, press the Enter key.
Resetting the settings:
Unless a user action (contact with the pressure-sensitive areas and keys) occurs within 30 seconds,
the system automatically exits the menu and the most recently confirmed setting defaults.
Heracell 150i / 240i GP 50159351_B_06-2020
control loop, the work space is not exposed to CO
2
77
gas
2