Preparation for Accurate Measurement (Executing Correction)
Checking Displaying/Setting Up Correction Data
Checking Displaying/Setting Up Correction Data
Checking displaying/setting up data for OPEN correction
Displaying data/Selecting parameter format for OPEN correction
For information on the procedure to display the data for OPEN correction, refer
to
Setting up data for OPEN correction
You can setup the data for OPEN correction by the following procedure.
Step 1.
Press the [Meas Setup] key.
Step 2.
Press the CORRECTION softkey.
Step 3.
Use the cursor keys to select the OPEN G-B field.
Step 4.
Select the G-B or Cp-G which you required.
Step 5.
Use the right cursor keys to select the A field.
Step 6.
Input your desired value for the primary parameter for OPEN.
Step 7.
Use the right cursor keys to select the B field.
Step 8.
Input your desired value for the secondary parameter for OPEN.
Checking displaying/setting up data for SHORT correction
Displaying data/Selecting parameter format for SHORT correction (6 digits)
For information on the procedure to display the data for SHORT correction,
refer to
Setting up data for SHORT correction
You can setup the data for SHORT correction by the following procedure.
Step 1.
Press the [Meas Setup] key.
Step 2.
Press the CORRECTION softkey.
Step 3.
Use the cursor keys to select the SHORT R-X field.
Step 4.
Select R-X or Ls-Rs which you required.
Step 5.
Use the right cursor keys to select the A field.
Step 6.
Input your desired value for the primary parameter for SHORT.
Step 7.
Use the right cursor keys to select the B field.
Step 8.
Input your desired value for the secondary parameter for SHORT.
110
"Checking data for OPEN correction" on page
"Checking data for SHORT correction" on page
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
46.