Servicing
Analog Output Testing
Analog Voltage Outputs
7-28
Model 48i Trace Level-Enhanced Instruction Manual
adjusted. Refer to the
follows.
8. Press
to reset the analog outputs to normal.
Figure 7–14. Rear Panel Analog Input and Output Pins
Table 7–4. Analog Output Channels and Rear Panel Pin Connections
Voltage Channel
Pin
1
14
2
33
3
15
4
34
5
17
6
36
Ground
16, 18, 19, 35, 37
"Analog Output
Calibration" procedure that
Current Outputs
Analog Voltage Inputs
Current Channel
1
2
3
4
5
6
Current Output
Return
Pin
15
17
19
21
23
25
16, 18, 20, 22, 24
Thermo Fisher Scientific