Thermo Scientific 48i Instruction Manual page 262

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

C-Link Protocol Commands
Measurement Configuration
B-30
Model 48i Trace Level-Enhanced Instruction Manual
Send:
set sample
Receive:
set sample ok
set zero
This command sets the zero/span valves to the zero mode. The example
below sets the instrument to zero mode that is, the instrument is reading
the sample gas.
Send:
set zero
Receive:
set zero ok
set span
This command sets the zero/span valves to the span mode. The example
below sets the instrument to span mode that is, the instrument is sampling
span gas.
Send:
set span
Receive:
set span ok
gas unit
This command reports the current gas units (ppm or mg/m
reports that the gas unit is set to ppm.
Send:
gas unit
Receive:
gas unit ppm
set gas unit
3
unit = | ppm | mg/m
|
This command sets the gas units to ppm or mg/m
the gas units to mg/m
3
Send:
set gas unit mg/m3
Receive:
set gas unit mg/m3 ok
pres comp
This command reports whether pressure compensation is on or off. The
example below shows that pressure compensation is on.
Send:
pres comp
Receive:
pres comp on
set pres comp onoff
These commands turn the pressure compensation on or off. The example
below turns pressure compensation off.
Send:
set pres comp off
Receive:
set pres comp off ok
.
3
). The example
3
. The example below sets
Thermo Fisher Scientific