Operation
Service Menu

Internal Set Temperature

Scrubber Test

3-76
Model 48i Trace Level-Enhanced Instruction Manual
INITIAL S/R RATIO:
CURRENTLY:
MEASURED:
SET TO:

CHANGE VALUE
RANGE
AVG
The Internal Set Temperature screen allows the user to set the operating
point for internal set temperature (for fan control).
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Internal Set Temp.
INTERNAL TEMP SETPOINT:
CURRENTLY:
SET TO:
RANGE
AVG
The Scrubber Test screen allows the user to initiate a scrubber efficiency
test, or to stop a test that is currently in progress. Typically, the efficiency
test should run for at least 10 minutes. When the efficiency test is
initiated, a timer is started and the efficiency test will automatically shut
off. The scrubber test results allow the user to view the current CO
reading, the span gas concentration, and the scrubber effectiveness,
expressed as a percent efficiency.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Scrubber Test.
1.079620
1.130966
1.131524

MOVE CURSOR
SAVE
DIAGS ALARM
o
43.0
C
o
43.5
C ?

CHANGE VALUE
SAVE VALUE
DIAGS ALARM
?