Operation
Service Menu
Coefficients
Default Coefficients
Preamp Board Calibration
3-78
Model 48i Trace Level-Enhanced Instruction Manual
SET CAL GAS POINT:
RANGE:
CAL FACTOR 1:
CAL GAS 1:

CHANGE VALUE
RANGE
AVG
The Coefficients screen allows the user to view and re-calculate the
calibration coefficients.
In the Main Menu, choose Service > Multipoint Cal > Choose Cal
Point > Coefficients.
COEFFICIENTS:
RANGE
AVG
The Default Coefficients screen allows the user to view and reset the
calibration coefficients to default values.
In the Main Menu, choose Service > Multipoint Cal > Choose Cal
Point > Coefficients.
COEFFICIENTS:
SET TO DEFAULTS
RANGE
AVG
The Pre-amp Board Calibration screen is used to adjust the preamp board
calibration parameters. The AGC circuitry is turned off while the preamp
10000
1.0000
0040.000

MOVE CURSOR
SAVE
DIAGS ALARM
1: 1.000000
2: 1.000000
3: 1.000000
RECALCULATE?
DIAGS ALARM
1: 1.000000
2: 1.000000
3: 1.000000
DIAGS ALARM
?