Operation
Instrument Controls Menu
Output Relay Settings
3-40
Model 48i Trace Level-Enhanced Instruction Manual
Note Analog Input Config is only displayed if the optional I/O Expansion
Board is installed.
In the Main Menu, choose Instrument Controls > I/O Configuration.
●
The Output Relay Settings menu displays a list of the 10 digital output
relays available, and allows the user to select the logic state or instrument
parameter for the relay selected.
In the Main Menu, choose Instrument Controls > I/O Configuration >
●
Output Relay Settings.
Logic State
The Logic State screen is used to change the I/O relay to either normally
open or normally closed.
Press
●
▲
I/O CONFIGURATION:
>OUTPUT RELAY SETTINGS
DIGITAL INPUT SETTINGS
ANALOG OUTPUT CONFIG
ANALOG INPUT CONFIG
RANGE
AVG
DIAGS ALARM
OUTPUT RELAY SETTINGS:
>1
NOP
2
NOP
3
NOP
4
NOP
5
NOP
6
NOP
7
NOP
RANGE
AVG
DIAGS ALARM
to toggle and set the logic state open or closed.
OUTPUT RELAY SETUP:
>LOGIC STATE
INSTRUMENT STATE
RANGE
AVG
DIAGS ALARM
CONC ALARM
NONE
UNITS
GEN ALARM
NONE
NONE
NONE
OPEN
Thermo Fisher Scientific