Thermo Scientific 48i Instruction Manual page 254

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

C-Link Protocol Commands
Datalogging
B-22
Model 48i Trace Level-Enhanced Instruction Manual
set copy lrec to sp
set copy srec to sp
set copy stream to sp
These commands copy the current contents of the long record, short
record, or streaming data list into the scratch pad (sp). These commands
are useful in easy modification of current long record, short record, or
streaming data lists. The example below copies the current list of long
records into the scratch pad.
Send:
set copy lrec to sp
Receive:
set copy lrec to sp ok
sp field number
This command reports the variable number and name stored at index in the
scratch pad list. The example below shows that the field 5 in the scratch
pad is set to index number 13, which is for the variable pressure.
Send:
sp field 5
Receive:
sp field 5 13 pres
set sp field number value
number = 1-32 is the maximum number of fields in long and short record
lists.
number = 1-18 is for streaming data lists.
This command sets the scratch pad field number (item number in scratch
pad list) to value, where value is the index number of a variable in the
analog out variable list. Available variables and their corresponding index
numbers may be obtained using the command "list var aout". The "set sp
field" command is used to create a list of variables which can then be
transferred into the long record, short record, or streaming data lists, using
the "set copy sp to lrec", "set copy sp to srec", or "set copy sp to stream"
commands.
Send:
set sp field 1 34
Receive:
set sp field 1 34 ok
stream per
This command reports the currently set time interval in seconds for
streaming data.
Send:
stream per
Receive:
stream per 10
Thermo Fisher Scientific