Download Print this page

Thermo Scientific 48i Instruction Manual page 241

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Thermo Fisher Scientific
Selection
Averaging Time (seconds)
8
120
9
180
10
240
11
300
co
high co
low co
These commands report the measured CO concentration when operating
in single range, or high and low CO when operating in dual or auto range
mode. The example below shows that the CO concentration is 40 ppm.
Send:
co
Receive:
co 0040E+0 ppm
flow
This command reports the current sample flow. The example below reports
that the current sample flow is 1.108 liters/minute.
Send:
flow
Receive:
flow 1.108 1/m
chamber temp
This command reports the current optical chamber temperature. The
example below reports that the current optical chamber temperature is 45.2
°C.
Send:
chamber temp
Receive:
chamber temp 45.2 deg C
internal temp
This command reports the current internal instrument temperature. The
first reading is the temperature being used in instrument calculations. The
second temperature is the actual temperature being measured. If
temperature compensation is on, then both temperature readings are the
same. If temperature compensation is off, a temperature of 30 °C is used as
the default temperature even though the actual internal temperature is 27.2
°C. The example below shows that temperature compensation is on and
that the internal temperature is 27.2 °C.
Send:
internal temp
Receive:
internal temp 27.2 deg C, actual 27.2
C-Link Protocol Commands
Model 48i Trace Level-Enhanced Instruction Manual
Measurements
B-9