Operation
Alarms Menu
Min and Max AGC
Intensity Limits

Motor Speed

Zero and Span
3-66
Model 48i Trace Level-Enhanced Instruction Manual
AGC INTENSITY:
ACTUAL
>MIN
MAX
RANGE
The Minimum AGC Intensity alarm limit screen is used to change the
minimum sample flow alarm limit. The minimum and maximum AGC
Intensity screens function the same way.
In the Main Menu, choose Alarms > AGC Intensity > Min or Max.
AGC INTENSITY:
ACTUAL MIN:
SET MIN TO:
RANGE
The Motor Speed screen (read only) displays the current motor speed. A
reading other than 100.0% indicates a problem with either the motor or
the power supply.
In the Main Menu, choose Alarms > Motor Speed.
MOTOR SPEED:
ACTUAL
RANGE
The Zero Span Check screen allows the user to view the status of the most
recent zero check and set the maximum zero check offset.
Check
196646 Hz
150000 Hz
300000 Hz
AVG
DIAGS ALARM
150000 Hz
180000 Hz ?

INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
100.0 %
AVG
DIAGS ALARM