Figure 7-15. Rear Panel Analog Input And Output Pins - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Thermo Fisher Scientific
Equipment Required:
Multimeter
Use the following procedure to test the analog outputs.
1. Connect a meter to the channel to be tested.
analog output pins and
Analog Voltage Outputs
Figure 7–15. Rear Panel Analog Input and Output Pins
2. From the Main Menu, press
The Test Analog Outputs screen appears.
3. Press
panel terminal pins where the meter is connected, and press
The Set Analog Outputs screen appears.
4. Press
The Output Set To line displays Zero.
Table 7–4
>
to scroll to Test Analog Outputs, and press
to scroll to the desired channel corresponding to the rear
to set the output to zero.
Analog Output Testing
Figure 7–15
identifies the associated channels.
Current Outputs
Analog Voltage Inputs
to scroll to Diagnostics, > press
Model 42i Instruction Manual
Servicing
shows the
.
.
7-27