Temperature Calibration
Analog Output
Calibration
Thermo Fisher Scientific
board is being calibrated. The value should be set so that the sample and
reference values are around 150000 Hz.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Preamp Calibration.
PREAMP BOARD CAL:
SAMPLE:
REFERENCE:
SET TO:
RANGE
AVG
The Temperature calibration screen allows the user to view and set the
ambient temperature sensor calibration. The temperature calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Temperature Calibration.
CALIBRATE AMBIENT TEMP:
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
The Analog Output Calibration menu is a selection of 6 voltage channels
and 6 current channels (if I/O expansion board option is installed) to
calibrate, and allows the user to select the calibration action zero or span.
The analog output calibration is visible only when the instrument is in
127807 Hz
148256 Hz
51

CHANGE VALUE
SAVE VALUE
DIAGS ALARM
"Service
Mode" earlier in the chapter.
o
32.3
C
o
032.5
C ?

MOVE CURSOR
SAVE
DIAGS ALARM
Model 48i Trace Level-Enhanced Instruction Manual
Operation
Service Menu
?
3-79