C-Link Protocol Commands

Datalogging

Datalogging
B-16
Model 48i Trace Level-Enhanced Instruction Manual
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
clr records
This command will clear all long and short records that have been saved.
Send:
clear records
Receive:
clear records ok
set clr lrecs
set clr srecs
These commands will clear only the long records or only the short records
that have been saved. The example below clears short records.
Send:
set clr srecs
Receive:
set clr srecs ok
data treatment lrec
data treatment srec
These commands report the current selection of data treatment for
concentrations in the long records (lrecs) or short records (srecs). The
example below reports the data treatment for concentrations in lrec is
minimum.
Send:
data treatment lrec
Receive:
data treatment lrec min
set data treatment lrec string
set data treatment srec string
string = | cur | avg | min | max |
These commands set the data treatment to current, average, minimum, or
maximum for the concentration values recorded in the long records (lrecs)
or short records (srecs). The example below sets the data treatment for
concentrations in lrec to minimum.
Send:
set data treatment lrec min
Receive:
set data treatment lrec min ok
list din
list dout