Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Thermo Fisher Scientific
co bkg
This command reports the current CO backgrounds. The example below
reports that the CO background is 1.4 ppm.
Send:
co bkg
Receive:
co bkg 1.4 ppm
set co bkg value
This command is used to set CO backgrounds to user-defined values to
value, where value is a floating-point representation of the background in
current selected units. The example below sets the CO background to 1.4
ppm.
Send:
set co bkg 1.400
Receive:
set co bkg 1.400 ok
init ratio
This command reports the initial sample/reference ratio. The example
below reports that the initial ratio was 1.16210.
Send:
initial ratio
Receive:
initial ratio 1.16210
sp conc
high sp conc
low sp conc
These commands report span concentration in single range mode, or the
high and low span concentrations in dual or auto range mode. If the mode
is incorrect, the instrument responds with "can't, wrong settings". The
example below reports the span gas concentration in single range mode.
Send:
sp conc
Receive:
sp conc 1000
set sp conc value
set high sp conc value
set low sp conc value
These commands set the span concentrations to user-defined values to
value, where value is a floating-point representation of the span
concentration in current selected units. The example below sets the span
concentration to 1000 ppb in the single range mode.
Send:
set sp conc 1000
Receive:
set sp conc 1000 ok
C-Link Protocol Commands
Model 48i Trace Level-Enhanced Instruction Manual
Calibration
B-25