Operation
Alarms Menu
Min and Max Internal
Temperature Limits
Bench Temperature
3-62
Model 48i Trace Level-Enhanced Instruction Manual
In the Main Menu, choose Alarms > Internal Temp.
INTERNAL TEMPERATURE:
ACTUAL
>MIN
MAX
RANGE
AVG
The Minimum Internal Temperature alarm limit screen is used to change
the minimum internal temperature alarm limit. The minimum and
maximum internal temperature screens function the same way.
In the Main Menu, choose Alarms > Internal Temp > Min or Max.
INTERNAL TEMPERATURE:
ACTUAL MIN:
SET MIN TO:
RANGE
AVG
The Bench Temperature screen displays the current bench temperature and
sets the minimum and maximum alarm limits. The bench temperature is
regulated at 50.0 degrees C. Acceptable alarm limits range from 40 to 59
°C. If the bench temperature reading goes beyond either the minimum or
maximum limit, an alarm is activated. The word "ALARM" appears in the
Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Bench Temp.
BENCH TEMPERATURE:
ACTUAL
>MIN
MAX
RANGE
AVG
o
40.6
C
o
38.0
C
o
43.0
C
DIAGS ALARM
o
38.0
C
o
40.0
C ?
