Diagnostics

Thermo Fisher Scientific
set alarm trig conc value
These commands set the CO concentration alarm minimum value, where
value is set to either floor or ceiling, according to
below sets the CO concentration minimum alarm trigger to ceiling.
Send:
set alarm trig conc 1
Receive:
set alarm trig conc 1 ok
Table B–3. Alarm Trigger Values
Value
Alarm Trigger
00
Floor
01
Ceiling
agc int
This command reports the current intensity of the reference channel AGC
circuit. The example below reports that the current AGC intensity is
200,000 Hz.
Send:
agc int
Receive:
agc int 250000 Hz
bias voltage
This command reports the current IR bias supply voltage. The example
below reports that the bias voltage is -102.3 volts
Send:
bias voltage
Receive:
bias voltage -102.3 V
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, positive 18 IR, positive 18
MOT, and VBIAS. Each voltage value is separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2 17.9 17.9
C-Link Protocol Commands
Table
B–3. The example
Model 48i Trace Level-Enhanced Instruction Manual
Diagnostics
B-15