Figure 7-14. Rear Panel Analog Input And Output Pins - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Thermo Fisher Scientific
The Set Analog Outputs screen appears.
4. Press
The Output Set To line displays Zero.
5. Check that the meter is displaying the zero value. If the meter reading
differs by more than one percent of the full-scale output, the analog
outputs should be adjusted. Refer to the
procedure that follows.
6. Press
The Output Set To line displays Full-Scale.
7. Check that the meter is displaying a full-scale value. If the meter
reading differs by more than one percent of the full-scale output, the
analog outputs should be adjusted. Refer to the
Calibration" procedure that follows.
8. Press
Analog Voltage Outputs
Figure 7–14. Rear Panel Analog Input and Output Pins
to set the output to zero.
to set the output to full-scale.
to reset the analog outputs to normal.
Analog Output Testing
"Analog Output
Calibration"
"Analog Output
Current Outputs
Analog Voltage Inputs
Model 48i Instruction Manual
Servicing
7-29