Calibrate Co Coefficient; Zero/Span Check - Thermo Scientific 48i Instruction Manual

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Calibration Menu
Calibrate CO
Coefficient

Zero/Span Check

3-20
Model 48i Trace Level-Enhanced Instruction Manual
CO BACKGROUND:
CO:
CURRENTLY:
RANGE
AVG
The Calibrate CO Coefficient screen is used to adjust the CO coefficient
and enter the span concentration. The display shows the current CO
concentration reading. The next line of the display is where the CO
calibration gas concentration is entered.
The CO span coefficient is calculated, stored, and used to correct the
current CO reading. For more information about calibration, see Chapter
4, "Calibration". In dual or auto range modes, "HIGH" or "LOW" is
displayed to indicate the calibration of the high or low coefficient.
In the Main Menu, choose Calibration > Cal Coefficient.
CALIBRATE CO:
CO:
SPAN CONC:

CHANGE VALUE
RANGE
AVG
The Zero/Span Check menu is used to program the instrument to perform
fully automated zero and span checks or adjustments. Total Duration Hour
is the sum of zero, span, and purge duration minutes. Zero and Span
Calibration Reset are toggle items that change between yes or no when
selected, and displayed if auto calibration is installed.
In the Main Menu, choose Calibration > Zero/Span Check.
1.4
0.0 ?
SET CO TO ZERO
DIAGS ALARM
79.1
000080.0

MOVE CURSOR
SAVE
DIAGS ALARM
?