C-Link Protocol Commands
I/O Configuration
B-40
Model 48i Trace Level-Enhanced Instruction Manual
list var aout
list var dout
list var din
These commands report the list of index numbers, and the variables
(associated with that index number) available for selection in the current
mode (determined by single/dual/auto, gas mode) for analog output, digital
output and digital inputs. The index number is used to insert the variable
in a field location in a list using "set sp field index". The example below
reports the list of analog output, index numbers, and variables.
Send:
list var aout
Receive:
list var aout
index variable
0 none
1 co
5 sr
8 intt
9 cht
10 auxt
13 pres
14 smplfl
15 intensity
16 speed
28 biasv
relay stat
This command reports the current relay logic normally "open" or normally
"closed," if all the relays are set to same state, that is all open or all closed.
The example below shows that the status when all the relays logic is set to
normally "open".
Send:
relay stat
Receive:
relay stat open
Note If individual relays have been assigned different logic then the
response would be a 4-digit hexadecimal string with the least significant
byte (LSB) being relay no 1.