C-Link Protocol Commands
Calibration
B-24
Model 48i Trace Level-Enhanced Instruction Manual
Send:
set cal low co coef
Receive:
set cal low co coef ok
co coef
high co coef
low co coef
These commands report CO coefficients in single range mode, or the high
and low range coefficients in dual or auto range mode. If the mode is
incorrect, the instrument responds with "can't, wrong settings". The
example below reports that the CO coefficient is 1.200.
Send:
co coef
Receive:
co coef 1.200
set co coef value
set high co coef value
set low co coef value
These commands set the CO coefficients to user-defined values to value,
where value is a floating-point representation of the coefficient. The
example below sets the CO coefficient to 1.200.
Send:
set co coef 1.200
Receive:
set co coef 1.200 ok
coef 0
coef 1
coef 2
high coef 0
high coef 1
high coef 2
low coef 0
low coef 1
low coef 2
The "coef 0", coef 1", and coef 2" commands report the coefficients of the
curve developed from the Hi Multi-Point Calibration in single range mode.
The "high coef 0", high coef 1", and high coef 2" commands report the
coefficients of the curve developed from the Hi Multi-Point Calibration in
dual or auto range mode. The "low coef 0", low coef 1", and low coef 2"
commands report the coefficients of the curve developed from the Lo
Multi-Point Calibration in dual and auto range mode. The example below
reports the coefficient 0 having a value of 1.005.
Send:
coef 0
Receive:
coef 0 1.005 ok
Thermo Fisher Scientific