Preventive Maintenance

Replacement Parts

Replacement Parts
Cleaning the
Outside Case
Cleaning the
5-2
Model 48i Trace Level-Enhanced Instruction Manual
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component. For more information
about appropriate safety precautions, see the "Servicing" chapter.
See the "Servicing" chapter for a list of replacement parts and the associated
replacement procedures.
WARNING If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired
Do not attempt to lift the instrument by the cover or other external
fittings.
Clean the outside case using a damp cloth being careful not to damage the
labels on the case.
Equipment Damage Do not use solvents or other cleaning products to
clean the outside case.
Best results are obtained when the optics are cleaned prior to calibration.
The cleanliness of the mirrors should be checked any time the AGC
Optics
intensity is below 200000 Hz, since one cause of low output is light
attenuation due to dirt on the mirrors.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component. For more information
about appropriate safety precautions, see the "Servicing" chapter.
Use the following procedure to clean the mirrors.
2. Turn off power and disconnect power line.