Pressure Compensation - Thermo Scientific 48i Instruction Manual

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Compensation
Screen Contrast
Thermo Fisher Scientific
Pressure
Pressure compensation provides compensation for any changes to the
instrument's output signal due to reaction chamber pressure variations. The
effects of reaction chamber pressure changes on the analyzer's subsystems
and output have been empirically determined. This empirical data is used
to compensate for any change in reaction chamber pressure.
When pressure compensation is on, the first line of the display represents
the current pressure in the reaction chamber. When pressure compensation
is off, the first line of the display shows the factory standard pressure of 760
mmHg.
In the Main Menu, choose Instrument Controls > Pressure
Compensation.
The Screen Contrast screen is used to change the contrast of the display.
Values between 0 and 100% in increments of 10 are available. Changing
the screen contrast may be necessary if the instrument is operated at
extreme temperatures.
In the Main Menu, choose Instrument Controls > Screen Contrast.
TEMPERATURE COMPENSATION:
COMP TEMP:
CURRENTLY:
SET TO:
TOGGLE VALUE
RANGE
AVG
DIAGS ALARM

PRESSURE COMPENSATION:

COMP PRES:
CURRENTLY:
SET TO:
TOGGLE VALUE
RANGE
AVG
DIAGS ALARM
o
32.6
C
ON
OFF ?
743.0 mmHg
ON
OFF ?
Model 48i Trace Level-Enhanced Instruction Manual
Operation
Instrument Controls Menu
3-51