Analog Output Calibrate
Analog Input
Calibration
Thermo Fisher Scientific
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
Full-Scale
a meter to the output and adjust output until it reads the value shown in
the set output to: field.
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel > Calibrate Full Scale.
The Analog Input Calibration menu is a selection of 8 analog inputs to
calibrate, and allows the user to select the calibration action zero or span.
The analog input calibration is visible only when the instrument is in
service mode. For more information on the service mode, see
Mode" earlier in the chapter.
Note This screen is present only when the I/O expansion board is
installed.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Analog Input Calibration.
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
AVG
DIAGS ALARM
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
INPUT CHANNEL 6
INPUT CHANNEL 7
RANGE
AVG
DIAGS ALARM
SPAN
V1
6397

INC/DEC
10 V
Model 48i Trace Level-Enhanced Instruction Manual
Operation
Service Menu
"Service
3-81