Analog Input Configuration

Thermo Fisher Scientific
Concentrations
The Analog Input Configuration menu displays a list of the 8 analog input
channels available for configuration. This screen is only displayed if the I/O
expansion board option is installed. Configuration includes entering
descriptor, units, decimal places, choice of 2-10 points in the table, and
corresponding number of points selected.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config.
Descriptor
The Descriptor screen allows the user to enter the descriptor for the
selected analog input channel. The descriptor is used in datalogging and
streaming data to report what data is being sent out. The descriptor may be
from 1 to 3 characters in length, and defaults to IN1 to IN8 (user input
channel number).
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Descriptor.
Other Measurements
Motor Speed
Bias Supply
Scrubber Efficiency
ANALOG INPUT CONFIG:
>CHANNEL 1
CHANNEL 1
CHANNEL 1
CHANNEL 1
CHANNEL 1
CHANNEL 1
CHANNEL 1
RANGE
AVG
DIAGS ALARM
ANALOG INPUT 01 CONFIG:
>DESCRIPTOR
UNITS
DECIMAL PLACES
TABLE POINTS
POINT 1
POINT 2
RANGE
AVG
DIAGS ALARM
Analog Inputs
IN1
IN2
IN3
IN4
IN5
IN6
IN7
IN1
V
2
2
Model 48i Trace Level-Enhanced Instruction Manual
Operation
Instrument Controls Menu
3-47