Calibrate Pressure Span
Restore Default Pressure
Calibration
Flow Calibration
Thermo Fisher Scientific
The Calibrate Pressure Span screen allows the user to view and set the
pressure sensor calibration span point.
Note The plumbing going to the pressure sensor should be disconnected so
the sensor is reading ambient pressure before performing the span
calibration. The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before
calibrating.
In the Main Menu, choose Service > Pressure Calibration > Span.
●
The Restore Default Pressure Calibration screen allows the user to reset the
pressure calibration configuration values to factory defaults.
In the Main Menu, choose Service > Pressure Calibration > Set
●
Defaults.
RESTORE DEFAULT CAL:
RANGE
AVG
DIAGS ALARM
The Flow Calibration submenu is used to calibrate the flow sensor to zero,
span, or restore factory default values. The flow calibration is visible only
▲
CALIBRATE PRESSURE SPAN:
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
RANGE
AVG
DIAGS ALARM
RESTORE
753.0 mmHg
760.0
mmHg ?
SAVE
RESTORE DEFAULT CAL:
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
RANGE
AVG
Model 48i Trace Level-Enhanced Instruction Manual
Operation
Service Menu
RESTORE
TO?x