Operation
Service Menu
Analog Output Calibrate
3-80
Model 48i Trace Level-Enhanced Instruction Manual
service mode. For more information on the service mode, see
Mode" earlier in the chapter.
Note This adjustment should only be performed by an instrument service
technician.
▲
In the Main Menu, choose Service > Analog Out Calibration.
●
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
RANGE
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE
The Analog Output Calibrate Zero screen allows the user to calibrate the
Zero
zero state of the selected analog output. The operator must connect a meter
to the output and adjust the output until it reads 0.0 V on the meter.
In the Main Menu, choose Service > Analog Out Calibration > Select
●
Channel > Calibrate Zero.
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
AVG
DIAGS ALARM
AVG
DIAGS ALARM
ZERO
INC/DEC
0.0 V
AVG
DIAGS ALARM
"Service
V1
100
Thermo Fisher Scientific