Operation
Service Menu
Analog Output Calibrate
3-80
Model 48i Trace Level-Enhanced Instruction Manual
service mode. For more information on the service mode, see
Mode" earlier in the chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Analog Out Calibration.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
RANGE
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE
The Analog Output Calibrate Zero screen allows the user to calibrate the
Zero
zero state of the selected analog output. The operator must connect a meter
to the output and adjust the output until it reads 0.0 V on the meter.
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel > Calibrate Zero.
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
AVG
DIAGS ALARM
AVG
DIAGS ALARM
ZERO

INC/DEC
0.0 V
AVG
DIAGS ALARM
"Service
V1
100
Thermo Fisher Scientific