Non-Alarm - Thermo Scientific 48i Instruction Manual

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Instrument Controls Menu
Digital Input Settings
3-42
Model 48i Trace Level-Enhanced Instruction Manual

Non-Alarm

The Non-Alarm status screen allows the user to select the non-alarm status
for the selected relay output. The selected item is shown by "<--" after it.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > Select Relay > Instrument State > Non-Alarm.
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note The digital inputs must be asserted for at least one second for the
action to be activated.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings.
Logic State
The Logic State screen is used to change the I/O relay to either normally
open or normally closed. The default state is open, which indicates that a
relay connected between the digital input pin and ground is normally open
NON ALARM STATUS ITEMS:
>NONE
AUTORANGE
LOCAL/REMOTE
SERVICE
UNITS
ZERO MODE
SPAN MODE
RANGE
AVG
DIAGS ALARM
SAMPLE MODE
PURGE MODE
SCRUB TST MODE
DIGITAL INPUT SETTINGS:
>1
NOP
SET BACKGROUND
2
NOP CAL TO
3
NOP
AOUTS TO ZERO
4
NOP
AOUTS TO FS
5
NOP
6
NOP
7
NOP
RANGE
AVG
DIAGS ALARM
LO SPAN
NONE
NONE
NONE
Thermo Fisher Scientific