C-Link Protocol Commands
Datalogging
B-20
Model 48i Trace Level-Enhanced Instruction Manual
erec layout
lrec layout
srec layout
These commands report the layout (string indicating the data formats) for
data that is sent out in response to the erec, lrec, srec, and related
commands. For details on how to interpret the strings, see "Record Layout
Definition" later in this appendix.
Send:
lrec layout
Receive:
lrec layout %s %s %lx %f %f %f %f
t D L ffff
s/r pres intensity motor
lrec mem size
srec mem size
These commands report the long and short records that can be stored with
the current settings and the number of blocks reserved for long and short
records. To calculate the number of short records per block, add 2 to the
number of records, and then divide by the number of blocks. The example
below shows that 10 blocks were reserved for long records, and the
maximum number of long records that can be stored in memory is 2038.
Send:
lrec mem size
Receive:
lrec mem size 2038 recs, 10 blocks
lrec per
srec per
These commands report the long and short records logging period. The
example below shows that the short record logging period is 5 minutes.
Send:
srec per
Receive:
srec per 5 min
set srec per value
set srec per value
value = | 1 | 5 | 15 | 30 | 60 |
These commands set the long and short records logging period to value in
minutes. The example below sets the long record logging period to 15
minutes.
Send:
set lrec per 15
Receive:
set lrec per 15 ok
Thermo Fisher Scientific