Multi-Point Calibration

Calibrate Point 1/2/3

Thermo Fisher Scientific
SCRUBBER EFFICIENCY:
TEST PHASE: SCRUBBER
TEST GAS CONC:
CURRENT CONC:
% EFFICIENCY:
RANGE
AVG
DIAGS ALARM
Up to three gas concentrations (cal-points) for each range may be calibrated
to using the following steps. Three cal-points will give the most accurate
readings over the entire range. The calibration process is sequential and will
work properly if all steps are followed in order. The example below shows
the multi-point calibration screen in single range mode. In dual or auto
range modes, "HI" or "LO" is displayed to indicate the calibration of the
high or low concentrations and function the same way.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Multipoint Calibration.
The Calibrate Point 1 screen allows the user to view and set the selected
calibration point. The calibrate 2 and calibrate 3 screens function the same
way.
In the Main Menu, choose Service > Multipoint Cal > Calibrate 1, 2
or 3.
0.000
0.000
0.0%
START
MULTI-POINT CAL:
>CALIBRATE POINT 1
CALIBRATE POINT 2
CALIBRATE POINT 3
COEFFICIENT

DEFAULT COEFFICIENTS

RANGE
AVG
DIAGS ALARM
SCRUBBER EFFICIENCY:
TEST PHASE: SCRUBBER
TEST GAS CONC:
CURRENT CONC:
% EFFICIENCY:
PHASE TIME: 599s
RANGE
AVG
DIAGS ALARM
Model 48i Trace Level-Enhanced Instruction Manual
Operation
Service Menu
0.000
0.000
0.0%
ABORT
3-77