Download Print this page

Thermo Scientific 48i Instruction Manual page 250

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

C-Link Protocol Commands
Datalogging
B-18
Model 48i Trace Level-Enhanced Instruction Manual
Send:
lr01
Receive:
lr01
10:15 05-12-03 flags 9c040000 co 7349E+0 loco 5994E+0
intt 33.2 cht 44.7 pres 758.9 smplfl 1.085 speed 100.0
biasv -115.5 intensity 1999940
erec
This command returns a brief description of the main operating conditions
at the time the command is issued (i.e. dynamic data). The example below
shows a typical response. The format is defined by the current settings of
"format" and "erec format" commands. For details on how to decode the
flag fields within these records, see the "flags" command.
Send:
erec
Receive:
erec
09:48 04-06-05 flags 9C040510 co 0.000 4 loco -0.002
4 s/r 0.000 los/r 0.902 biasv -112.668 agci 96.500 intt
34.023 cht 47.995 smpfl 0.000 pres 0.000 avgt1 10 avgt2
10 cobkg -0.000 cocoef 1.000 lococoef 1.000 corange
10000000.000 locorange 10000000.000 motor 100.000
lrec
srec
lrec xxxx yy
srec xxxx yy
lrec aa:bb oo-pp-qq yy
srec aa:bb oo-pp-qq yy
xxxx = the number of past records
yy = the number of records to return (1 to 10)
aa = hours (01 to 24)
bb = minutes (01 to 59)
oo = month (01 to 12)
pp = day (01 to 31)
qq = year
These commands output long or short records. The output format is
determined by the "set lrec format" and "set srec format" commands. The
logging time is determined by the "set lrec per" and "set srec per"
commands.
In the following example, there are 740 long records currently stored in
memory. When the command lrec 100 5 is sent, the instrument counts
back 100 records from the last record collected (record 740), and then
returns 5 records: 640, 641, 642, 643, and 644. For details on how to
decode the flag fields within these records, see the "flags" command.
Thermo Fisher Scientific

Advertisement

loading