C-Link Protocol Commands
Communications Configuration
B-36
Model 48i Trace Level-Enhanced Instruction Manual
Send:
instr name
Receive:
instr name
CO Analyzer
CO Analyzer
instrument id
This command reports the instrument id.
Send:
instrument id
Receive:
instrument id 12
set instrument id value
This command sets the instrument id to value, where value is a decimal
number between 0 and 127 inclusive.
Note sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands. 5
Send:
set instrument id 12
Receive:
set instrument id 12 ok
mode
This command reports what operating mode the instrument is in: local,
service, or remote. The example below shows that the instrument is in the
remote mode.
Send:
mode
Receive:
mode remote
set mode local
set mode remote
These commands set the instrument to local or remote mode. The example
below sets the instrument to the local mode.
Send:
set mode local
Receive:
set mode local ok
program no
This command reports the analyzer's model information and program
version number, which will be dependant on the current version.
Send:
program no
Receive:
program no iSeries 48iTLE 01.02.12.095
set layout ack
This command disables the stale layout/layout change indicator ('*') that is
attached to each response if the layout has changed.