The Minimum Concentration alarm limit screen is used to change the
minimum concentration alarm limits. The minimum and maximum
concentration alarm limit screens function the same way.
In the Main Menu, choose Alarms > Concentration > Min or Max.
Min Trigger
The Minimum Trigger screen allows the user to view and set the
concentration alarm trigger type to either floor or ceiling. The minimum
alarm may be programmed as a floor trigger (alarm is triggered when the
concentration falls below the minimum value) or a ceiling trigger (alarm is
triggered when the concentration goes above the minimum value).
In the Main Menu, choose Alarms > Select Concentration >
MinTrigger.
CONCENTRATION:
ACTUAL
>MIN
MAX
MIN TRIGGER
RANGE
AVG
DIAGS ALARM
CONCENTRATION:
ACTUAL MIN:
SET MIN TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
MIN TRIG(CEILING/FLOOR):
ACTUAL TRIGGER:
SET TRIGGER TO:
TOGGLE AND SAVE VALUE
RANGE
AVG
DIAGS ALARM
58.3
1000
1000
CEILING
0.0
00010.00 ?
SAVE
CEILING
FLOOR ?