Figure 7-16. Rear Panel Analog Input And Output Pins - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Analog Voltage Outputs
Thermo Fisher Scientific
Figure 7–16. Rear Panel Analog Input and Output Pins
2. From the Main Menu, press
>
to scroll to Test Analog Outputs, and press
The Test Analog Outputs screen appears.
3. Press
to scroll to the desired channel corresponding to the rear
panel terminal pins where the meter is connected, and press
The Set Analog Outputs screen appears.
4. Press
to set the output to zero.
The Output Set To line displays Zero.
5. Check that the meter is displaying the zero value. If the meter reading
differs by more than one percent, the analog outputs should be
adjusted. Refer to the
follows.
6. Press
to set the output to full-scale.
The Output Set To line displays Full-Scale.
Current Outputs
Analog Voltage Inputs
to scroll to Diagnostics, > press
"Analog Output
Calibration" procedure that
Model 42i Trace Level Instruction Manual
Servicing
Analog Output Testing
.
.
7-31