Table B-5. Stream Time Values; Calibration - Thermo Scientific 48i Instruction Manual

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Calibration

Thermo Fisher Scientific
set stream per number value
number value = | 1 | 2 | 5 | 10 | 20 | 30 | 60 | 90 |120 | 180 | 240 | 300 |
This command sets the time interval between two consecutive streaming
data strings to number value in seconds. The example below sets the
number value to 10 seconds.
Send:
set stream per 10
Receive:
set stream per 10 ok
stream time
This command reports if the streaming data string will have a time stamp
attached to it or not, according to
Send:
stream time
Receive:
stream time 0
set stream time value
This command enables value, where value is to attach or disable time stamp
to streaming data string, according to
attaches a time stamp to streaming data.
Send:
set stream time 0
Receive:
set stream time 0 ok
Table B–5. Stream Time Values
Value
Stream Time
00
Attaches time stamp to streaming data string
01
Disables time stamp to streaming data string
set cal co bkg
This command will auto-calibrate the CO background. The example below
shows a successful auto-calibration of the CO background.
Send:
set cal co bkg
Receive:
set cal co bkg ok
set cal co coef
set cal high co coef
set cal low co coef
These commands will auto-calibrate CO coefficients based on CO span gas
concentrations. The high and low commands are only available in dual and
auto range mode. If the mode is incorrect, the instrument responds with
"can't, wrong settings". The example below shows a successful auto-
calibration of the low CO coefficient.
C-Link Protocol Commands
Table
B–5.
Table
B–5. The example below
Model 48i Trace Level-Enhanced Instruction Manual
Calibration
B-23