Operation
Instrument Controls Menu
Temperature
Compensation
3-50
Model 48i Trace Level-Enhanced Instruction Manual

User Value

The User Value screen allows the user to set the output value for the
corresponding input voltage for the selected table point in the conversion
table, from -9999999 to 99999999. The default table is a two-point table
with point 1: 0.00 V = 000.0 U and point 2: 10.00 V = 10.0 U.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Table Point > User Value.
Temperature compensation provides compensation for any changes to the
instrument's output signal due to internal instrument temperature
variations. The effects of internal instrument temperature changes on the
analyzer's subsystems and output have been empirically determined. This
empirical data is used to compensate for any changes in temperature.
When temperature compensation is on, the display shows the current
internal instrument temperature (measured by a thermistor on the bench).
When temperature compensation is off, the display shows the factory
standard temperature of 30 °C.
In the Main Menu, choose Instrument Controls > Temperature
Compensation.
TABLE POINT 01 VOLTS:
CURRENTLY:
SET TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
TABLE POINT 01 USER VAL:
CURRENTLY:
SET TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
0.00
00.00
SAVE
0.00
00000.00
SAVE