Operation
Instrument Controls Menu

Select Output Range

Minimum and Maximum
3-44
Model 48i Trace Level-Enhanced Instruction Manual
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config.
OUTPUT CHANNELS:
>ALL VOLTAGE CHANNELS
ALL CURRENT CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
RANGE
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
RANGE
The Select Output Range screen is used to select the hardware range for the
selected analog voltage output channel. Possible ranges for the voltage
outputs are: 0-100 mV, 0-1, 0-5, 0-10 V.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Select Channel > Select Range.
SELECT OUTPUT RANGE:
SELECTED OUTPUT:
CURRENTLY:

RANGE
The Minimum Value screen is used to edit the zero (0) to full-scale (100)
Value
value in percentages for the selected analog output channel. See
for a list of choices. In dual or auto range mode, "HI" or "LO" is displayed
to indicate high or low concentrations. Range Status is visible only in auto
range mode. The minimum and maximum output value screens function
the same way. The example below shows the set minimum value screen.
AVG
DIAGS ALARM
AVG
DIAGS ALARM
SET TO:
CHANGE VALUE
AVG
DIAGS ALARM
V1
0-10V
0-5V
?
SAVE
Thermo Fisher Scientific