Enable the GC/MS Interface and Oven

Procedure
1
2
3
4
5
5975/5977 Series MSD Operation Manual
Click Method > Instrument Setup > GC > Aux Heaters.
Select On for Thermal Aux 2.
Click Oven and select On.
Click Download Method to enable these temperature zones in the GC.
Save the method.
Operating in Electron Ionization (EI) Mode