Instrument Logbook Errors And Step By Step Repair Process; Fan Error (Main Fan Or Exhaust Fan); Initialization Error; Metering Sensor Error - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

Instrument logbook errors and step by step repair process

The errors in the instrument logbook can be classified in 8 groups. In this
section you can find a general step by step troubleshooting process for each of
them.

1. Fan error (Main fan or Exhaust fan)

2. Initialization error

3. Metering sensor error

1100 Series WS MWS Reference Manual
Make sure the fan is connected to the main board correctly
Examine the fan connector for irregularities. Correct if possible
Replace the defective fan (main fan: 3160-1017, exhaust fan: 3160-4097)
Replace the main board (G1367-69500)
Upgrade the firmware revision to A.04.14 or higher and the ChemStation
revision to A.08.04 or higher
Check the sample transport connectors on the main board/sample
transport
Check the connector on the bottom of the sample transport unit
Check for mechanical obstruction on the sample transport (X,Theta, Z)
Change sample transport unit (G1367-60019)
Change main board (G1367-69500)
Check the connectors on the SUD board
Check the sampling unit connector on the main board
Check the connector on the sampling unit
Change the analytical head (G1367-60003)
Change the sampling unit (G1367-60008)
Change the main board (G1367-69500)
Troubleshooting and Test Functions
3
99