B
Cabling Diagrams and Remote Start/Stop

Multi-instrument Cabling Examples

GC / ALS / Non-Agilent Data System

Number
1
2
35900-60670 APG remote
cable spade lug identification
Connector 1 
Signal name
9 pin (male)
1
GND
2
Prepare
3
Start
4
Shut down
5
Reserved
6
Power on
7
Ready
8
Stop
9
Start Request
Installation and First Startup
1. APG Remote Cable,
G1530-60930
Part number and description
G1530-60930, General use APG remote cable, 9-pin male/spade lug (0.5m)
G1530-60590, External event cable, 8-pin/spade lugs
Connector 2 
spade lugs
Black
White
Red
Green
Brown
Blue
Orange
Yellow
Violet
GC/ALS
2. External event cable,
8 pin/spade lugs, G1530-60590
G1530-60590 External event
cable spade lug identification
Pin
Color
1
Yellow
2
Black
3
Red
4
White
5
Orange
6
Green
7
Brown
8
Blue
9
Signal
24 V Out
24 V Out 2
Ground
Ground
Contact 1
Contact 1
Contact 2
Contact 2