Temperature Sensor Resistance; Ceramic And Glass Fuses - Agilent Technologies 7820A Service Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

Temperature sensor resistance

°C
+0°
+10°
100.00 103.90
100° 138.50 142.28
200° 175.84 179.51
300° 212.03 215.58
400° 247.06 250.50

Ceramic and glass fuses

Agilent 7820 GC Service
263
The approximate resistance of a temperature sensor is:
R = 100 + (.35  t)
Where R is resistance in ohms and t is sensor temperature in °C.
Temperature Sensor Resistance by Heater Temperature
+20°
+30°
+40°
107.79 111.67 115.54 119.40
146.06 149.82 153.57 157.32
183.17 186.82 190.46 194.08
219.13 222.66 226.18 229.69
253.34 257.34 260.75 264.14
There are two sets of fuses on the AC power board:
Two ceramic fuses protect the board from the oven heating
circuit (F3, F4)
Two glass fuses protect the board from the main supply (F1,
F2)
See the table below for the part numbers. When replacing both
the glass and ceramic fuse types, always replace them in pairs.
I.D
Description
.
+50°
+60°
+70°
123.24
127.07
161.04
164.76
197.70
201.30
233.19
236.67
267.52
270.89
System
+80°
+90°
+100°
130.89 134.70 138.50
168.47 172.16 175.84
204.88 208.46 212.03
240.15 243.61 247.06
274.25 277.60 280.93
Part
number

Advertisement

Table of Contents
loading