Holmium Oxide Test Failed; Illegal Temperature Value From Sensor On Main Board - Agilent Technologies 1260 Infinity G1315C User Manual

Diode array and multiple wavelength detector
Table of Contents

Advertisement

7
Error Information
Detector Error Messages

Holmium Oxide Test Failed

Probable cause
1
2
3
4
5

Illegal Temperature Value from Sensor on Main Board

Error ID: 1071
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable cause
1
2
138
Lamps switched off.
Defective or dirty flow cell.
Defective filter assembly.
Defective achromat assembly.
Defective PDA/optical unit.
Defective sensor or main board.
Detector is exposed to illegal ambient
conditions.
Suggested actions
Ensure the lamps are switched on.
Ensure the flow cell is inserted correctly, and is
free from contamination (cell windows, buffers
etc.).
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Verify that the ambient conditions are within
the allowed range.
Agilent 1260 Infinity DAD and MWD User Manual