Column Thermostat Calibration Problems

If the temperature cannot be calibrated, check the following:
• Has the thermostat front cover been closed correctly?
• Is the measuring device functioning correctly, and is calibrated according to
Hardware Failures
Probable hardware failures that can lead to a failed calibration procedure are:
• Defective or wrongly calibrated measuring device.
• Defective heater assembly.
• Defective ambient-temperature sensor.
• Defective CCM board.

Installing the Temperature Sensor

Installation of the temperature sensor is required for the temperature
calibration and temperature verification procedures.
The figures below refer to a specific type of temperature sensor (Heraeus, Quat340, quartz
N O T E
surface-temperature measurement sensor). Other sensors may require a different fixing.