2.2.5.3 Transfer Points with Stabilization Wait Timer

Press Transfer Points on the Menus screen. The Transfer Points screen appears
(Figure 2-9). Use this screen to set the Stabilization Wait.
Figure 2-9 Transfer Points with Stabilization Wait Timer
Stabilization Wait
2-28
VS Series Component Leak Detector
This feature is available February 2011.
NOTE
This appears on the Home screen after a successful startup. This
timer is necessary for the spectrometer stabilization and degassing.
The factory default is three (3) minutes. It is highly recommended, if
you power off the leak detector overnight, over a weekend, or
expose the leak detector to aggressive applications, to select the 9
or 15 minute selection.